Proline Promass 300 EtherNet/IP
Commissioning
15
4. Under
Local Area Network Settings
click the
LAN Settings
button.
This opens a new window with the
Local Area Network Settings
.
A0024311
5. Deselect the
Use a Proxy Server for Your LAN
checkbox and then click
Ok
to confirm.
4.2
Prerequisites - measuring device
4.2.1
Enabling the Web server
The Web server must be enabled in the measuring device (factory setting).
If the Web server is disabled it be enabled again via the
Web server functionality
parameter (→ 23). To do so, users can choose from the following operation options:
• Local display
• Operating tool e.g. FieldCare, DeviceCare, AMS Device Manager , SIMATIC PDM
4.2.2
Determining the IP address of the measuring device
The IP address of the measuring device is required to establish communication between
the measuring device (Web server) and a computer (client). The IP address used depends
on the setting of the DIP switches or on the system settings.
The IP address can be assigned to the measuring device in a variety of ways:
• Dynamic Host Configuration Protocol (DHCP), factory setting:
The IP address is automatically assigned to the measuring device by the automation
system (DHCP server).
• Hardware addressing:
The IP address is set via DIP switches
• Software addressing:
The IP address is entered via the
IP address
parameter (→ 23)
• DIP switch for "Default Ethernet Network Settings":
For establishing the network connection via the service interface (CDI-RJ45). The fixed
IP address 192.168.1.212 is used .
Summary of Contents for Proline Promass 300 EtherNet/IP
Page 35: ......