Proline Promass 300 EtherNet/IP
Commissioning
21
4.5
Setting the IP address
The IP address of the measuring device is required to establish communication between
the measuring device (Web server) and a computer (client). There are various options to
assign or specify the IP address depending on the hardware and software settings.
Assign or specify the IP
address via:
Description
DHCP (Dynamic Host
Configuration
Protocol)
• The measuring device is automatically assigned the IP address by the
automation system.
Hardware addressing
• The measuring device uses the IP address configured via the "IP address setting"
DIP switches.
• "IP address setting" DIP switch =
ON/OFF
, depending on the address
Software addressing
The measuring device uses the IP address configured in the
IP address
parameter
Use of the DIP switch:
Default Ethernet
Network Settings
• The measuring device uses the fixed IP address: 192.168.1.212
• DIP switch: default Ethernet network settings =
ON
• The device is connected via the CDI-RJ45 service interface.
• Following a restart, the measuring device can be connected via the CDI-RJ45
service interface or via the Ethernet network.
• To avoid conflict between IP addresses, the DIP switch should never be used
simultaneously on two measuring devices in the same Ethernet network.
1)
Factory setting
2)
For a temporary connection when servicing, for example, or if the IP address is not known. The measuring
device is disconnected from the network/automation system.
4.6
Overview of the Web server parameters
4.6.1
Language
Navigation
"Operation" menu → Web server language