Proline Promass 300 EtherNet/IP
Commissioning
9
4
Commissioning
Establishing a connection to the integrated Web server
1. Configure the computer → 9.
2. Check the settings on the measuring device and change them if necessary → 15.
3. Connect the measuring device to the computer .
4. Establish a connection to the Web server → 19.
5. Start the Web browser and access the operating menu → 20.
The measuring device can be operated via the Web server.
4.1
Prerequisites - computer
4.1.1
Hardware
Hardware
Interface
CDI-RJ45
WLAN
Interface
The computer must have an RJ45
interface.
The operating unit must have a
WLAN interface.
Connection
Standard Ethernet cable with RJ45
connector.
Connection via Wireless LAN.
Screen
Recommended size: ≥12" (depends on the screen resolution)
4.1.2
Software
Software
Interface
CDI-RJ45
WLAN
Recommended operating
systems
• Microsoft Windows 7 or higher.
• Mobile operating systems:
– iOS
– Android
Microsoft Windows XP is supported.
Web browsers supported
• Microsoft Internet Explorer 8 or higher
• Microsoft Edge
• Mozilla Firefox
• Google Chrome
• Safari
4.1.3
Configuring the computer
User rights
Appropriate user rights (e.g. administrator rights) for TCP/IP and proxy server
settings are necessary (for adjusting the IP address, subnet mask etc.).
Proxy server settings of the
Web browser
The Web browser setting
Use a Proxy Server for Your LAN
must be
deselected
.