6. LSServer, Scanner Shell and Download
2004-11
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
6-39
Optics
Optics
group box
The appearance of the
Optics
group box is dependent on whether dual scan,
single scan or dual scan sequential is selected. In dual scan sequential
Channel
1 1
st
Run
and
Channel 1 2
nd
Run
will appear instead of
Laser Channel 1
and
Laser Channel 2
.
Laser Channel 1
combo box
Select the appropriate laser from the
Laser Channel 1
combo box for excitation.
The selection of lasers is dependent on the type of instrument connected.
Laser Channel 2
combo box
Select the appropriate laser from the
Laser Channel 2
combo box for excitation.
The selection of lasers is dependent on the type of instrument connected.
Laser Channel 2 is only activated, if
Dual Parallel Scan Mode
has been
selected in the
General tab
.
Filter Channel 1
combo box
Select the appropriate filter from the
Filter Channel 1
combo box to match with
the selected excitation wavelength in Laser Channel 1.
All filters that have been defined in the filter slide which is inserted into Filter
Channel 1, can be selected from the Filter Channel 1 combo box.
Filter Channel 2
combo box
Select the appropriate filter from the
Filter Channel 2
combo box to match with
the selected excitation wavelength in Laser Channel 2.
All filters that have been defined in the filter slide which is inserted into Filter
Channel 2, can be selected from the Filter Channel 2 combo box.
Filter Channel 2 is only activated, if
Dual Parallel Scan Mode
has been selected
in the
General tab
.
PMT Gain Channel 1
edit box
Set
PMT Gain
to a value between 70 and 255 to regulate the detector sensitivity
in Channel 1.
The PMT Gain can also be adjusted online during the measurement via the
Status control
dialog box window. The
Status
dialog box is displayed after
clicking the
Start
button.
PMT Gain Channel 2
edit box
Set
PMT Gain
to a value between 70 and 255 to regulate the detector sensitivity
in Channel 2.
The PMT Gain can also be adjusted online during the measurement via the
Status control
dialog box window. The
Status
dialog box is displayed after
clicking the
Start
button.
PMT Gain Channel 2 (“
Channel 1 2
nd
Run”
when
Dual Scan Sequential
is
selected) is only activated, if the instrument is a
Dual Channel Instrument
and if
Dual Parallel Scan Mode
has been selected in the
General tab
.
Unusable filter laser excitations are signaled by red exclamation marks.