6. LSServer, Scanner Shell and Download
6-76
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
2004-11
The
Test Command
dialog box contains the following elements:
Command
combo box
Use the
Command
combo box edit field to type in the
command that should be sent to the instrument.
Each command entered is automatically added to the
Command
combo list: All commands sent can be
repeated by clicking on the corresponding list item.
Answer
edit box
The
Answer
edit box shows answers received from
connected instrument.
Empty command input
field after sending
check box
Selecting this box deletes commands typed in
Command
combo box after they have been sent to
the instrument. New commands can be entered
immediately.
If the same command needs to be repeated several
times this box should not be selected.
Send
button
Click the
Send
button to send the command, which is
in
Command
combo box edit field, to the connected
instrument.
Cancel
button
Click the
Cancel
button to close the dialog box.
Demo Mode Allowed
To enable the
Demo Mode
option button
in the
Setup Port
dialog box
,
select
Demo Mode Allowed
from the
Service
menu.
In
Demo Mode
the
Tecan LSServer
can be simulated, no instrument needs to
be connected. This option is used for training purposes.
For further information refer to the chapter 6.2 Tecan LS Scanner Interface
description in this manual.
Transport Lock…
To start the
Transport Lock wizard,
select
Transport Lock…
from the Service
menu. A wizard guides the user through the transport lock procedure..
The
Transport Lock
option is implemented in
Scanner Shell
as a wizard. It
represents workflow modules, which define a sequential procedure and guide the
user step-by-step to perform a specific complex activity.
For transportation purposes it is recommended to lock the connected instrument.
A transportation shock absorber pad is inserted between the plate transport and
the optic chassis to avoid damage to the optical parts.
For further information refer to the chapter 6.4.9 The Transport Lock Wizard
description in this manual.
During the
Transport Lock
procedure:
1. The instrument door is opened.
2. The plate transport is moved into the correct position.
3. The user is asked to insert the transportation shock absorber pad.
4. Instrument is locked.
5. The instrument door is closed.
As long as the instrument is locked, the
LSServer Measurement Control
cannot
be started. Thus scanning is not possible.
For information unlocking instrument see 6.4.10 The Transport Unlock Wizard
description in this manual.