6. LSServer, Scanner Shell and Download
6-30
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
2004-11
Scan Mode
Scan Mode
group box
In the
Scan Mode
group box, several scan modes can be selected.
•
Single
option button:
Selects the Single Scan Mode. One excitation (= laser) and one
emission (= filter) wavelength can be selected. The detection will be
performed in channel 1. This is appropriate for single label
experiments.
•
Dual Sequential
option button:
Selects the Dual Sequential Scan Mode. Two excitation and emission
wavelengths can be selected. This is appropriate for dual label
experiments in single channel instruments. The second label is
measured immediately after the first and the two emissions are
detected sequentially in channel 1.
•
Dual Parallel
option button:
Selects Dual Parallel Scan Mode. Two excitation and emission
wavelengths can be selected and are measured simultaneously. This
requires dual channel instruments as detection is performed in both
channels at once and scanning is done in half the time. The Dual
Parallel Scan Mode needs a Crosstalk correction (see Scan Settings
tab) and a valid set of K-factors in order to compensate for spectral
crosstalk.
For further information, refer to the chapter 6.4.6 Service Menu -
Crosstalk… description in this manual.
Stacker
Stacker
group box
•
Use Stacker
check box:
The
Use Stacker
check box will be available, if a stacker is connected
to the instrument and the instrument is turned on.
Selecting
Use Stacker
operates the instrument in an automatic mode.
Autosave will be enabled automatically.
For further information about
Stacker
and
Stacker Setup
refer to the
chapter 6.4.6 Service Menu -
Stacker Setup… description in this manual.
See also Tecan Connect Configuration.
Plate
Plate
group box
•
Move plate out after measurement
check box:
Selecting
Move plate out after measurement
opens the door and
moves out the plate transport after the measurement is finished.