409-35001
Rev C
70
of 92
Load Cells Panel
The "Load Cells" panel (see Figure 75) contains two bar graphs that show the individual force readings for
each load cell. This panel also contains the "Tare Load Cells" button which can be used to "zero" the load cells.
Press the "Tare Load Cells" button if the machine is reading a force value when no load is present. The "Load
Cell Tare Values" are displayed below the "Tare Load Cells" button. These values indicate the current offset
from the raw force reading that is used to correctly set the zero point for each load cell.
Figure 75
Calibration Panel
The Calibration panel allows access to the height calibration functions (see Figure 76) and the automatic load
cell (force) calibration function (see Figure 77) as well as showing information on the most recent calibration
performed (see Figure 79).
NOTE
TE Connectivity also offers a load cell calibration service.
The height calibration functions are located under the “Height” subpanel in the Calibration Panel.
Figure 76