TDI-Dynaload Division
WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
402828, Rev. B1
Page 57
Detail Language Elements
This is an alphanumerical listing that gives the syntax and required parameters for all elements in
the programmable load’ s syntax. The syntax is generic for all load ratings.
The following section provides the following information below for each command:
DESCRIPTION: Describes the function of the command and gives information about its
use.
COMMAND SYNTAX: Shows the proper form(s) of the command. Spaces are optional
as the command processor ignores these.
EXAMPLE: Gives an example of command usage.
QUERY SYNTAX: Provides the proper query form.
RETURNED PARAMETERS: Indicates the response format(s) to the query form of the
command. Where the TEXT ON and TEXT OFF modes responses differ, both will be
indicated.
AV1 Amps/Volt Pulsing Base Level
Description
This command is used to set the base level when pulse loading in the
Amps/Volt mode.
NOTE:
AVL or AVH must have been previously set.
Command Syntax
AV1 <NR2>
Example
AV1 25.0
This will set the base Amps/Volt level to 25 amps/volt
Query Syntax
AV1?
Returned Parameters TEXT ON: <NR2> amps/v
TEXT OFF: <NR2>