WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
TDI-Dynaload Division
Page 64
402828, Rev. B1
ERR? Error Register
This query returns the contents of the Error register and then clears
its contents. The response is either numeric or a text description
depending on the TEXT ON/OFF setting.
The value in this register is a number between 0 and 255 and is the
sum of the individual bit weights as defined below.
Bit
Weight
Description
Text On Response
Message
7
128
—
—
6
64
—
—
5
32
Command not allowed
NOT ALLOWED
4
16
Command too long
TOO LONG
3
8
Numeric Error
NUMERIC
2
4
—
—
1
2
Out of Range
RANGE
0
1
Unrecognized Command UNRECOGNIZED
Description
The text response will return a string containing the Text On
Response Message as listed above for each bit that is set. These will
be separated by commas. If no bits are set, “ NO COMMAND
ERROR” will be returned.
Command Syntax
NONE
Example
N/A
Query Syntax
ERR?
Returned Parameters TEXT ON: <text response string>
TEXT OFF: <NR1>