WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
TDI-Dynaload Division
Page 84
402828, Rev. B1
UV Under Voltage
Description
This command is used to set the Programmable Under Voltage
threshold. When the input voltage drops below this threshold the
load will disconnect from the device under test and indicate an
Under Voltage fault. Unlike the hardware Under Voltage, the load
will not automatically reconnect when the Voltage rises above the
threshold. Setting the Programmable Under Voltage to zero will
disable it. The programmable Under Voltage operation is not
affected by the UV bit in the Shutdown (SDN) register.
Command Syntax
UV <NR2>
Example
UV 9.6
This will set the Programmable Under Voltage to 9.6 Volts. The
load will shutdown if the Voltage on its input terminals drops below
this threshold.
UV 0
This will disable the Programmable Under Voltage feature.
Query Syntax
UV?
Returned Parameters Text On: <NR2> volts
Text Off: <NR2>
V? Read Voltage
Description
This query returns the actual Voltage in Volts measured at the load’ s
Voltage sense terminals.
NOTE:
If remote Voltage sensing is not used, this voltage may differ
from the output voltage of the device. This is due to the Voltage
drops in the conductors between the load and the device under
test. Therefore remote voltage sensing is recommended when
high currents are involved.
Command Syntax
NONE
Example
N/A
Query Syntax
V?
Returned Parameters TEXT ON: <NR2> volts
TEXT OFF: <NR2>