[Initialize] button:
Clears any previous adjustment equation(s) from the Data Logger.
ɜ
[Send Settings] button:
Transmits the adjustment equation to the Data Logger.
ɜ
[Close] button:
Quits and closes the window.
Ȇ
We cannot guarantee that after carrying out adjustment the measuring accuracy
will improve for all measuring ranges.
Ȇ
By carrying out adjustment on an already adjusted device without fi rst clicking
the Initialize button, the new adjustment equation will be be used to further adjust
the previously adjusted result. It is important to remember that if adjustment takes
place without initializing the device, the adjustment will include the fi rst equation
as well as the second and an over adjustment will occur. This over adjustment
will be especially large when making a 2-point adjustment and the measurement
falls outside of the range of the two points entered.
When you are carrying out adjustments and formulating equations please take
care with the settings and take personal responsibility for the process.
(For details about the RTR-51/52 Temperature Adjustment Function. see pp.42)