51
RTR-53: Value Adjustment Function
This function enables the user to make adjustments to measured
values to correct inaccuracies that may occur when compared to a
standard or reference temperature measurement. By setting an
adjustment equation beforehand, the adjustment function will
automatically calculate to record and display the post-adjusted
measurement. It is possible to select from two methods of
adjustment: 1-point adjustment and 2-point adjustment.
Ȇ
When you are carrying out adjustments and formulating equations please take
care with the settings and take personal responsibility for the process.
ɡ
Humidity: 1-point/2-point Adjustment
Before : Enter the actual measurement before adjustment.
After : Enter the desired measurement after adjustment.
Ȇ
Entries are valid to one decimal point.
Ȇ
The input value for [Post-adjusted Reading] must be less than
Ⱦ
5
ɓ
in
comparison to the input value for [Current Reading].
ɡ
Connect the RTR-51/52 to the Computer and set
up for communication.
See pp.25-26 for more details about how to connect.
ɡ
Adjustment Settings
1.
In the Main Window, click on the Data Logger icon. To open the
[Adjustment] Window, click on [Settings] in the Menu Bar and then
select [Adjustment Settings] from the pull down menu.
2.
Check [Adjust this Channel], and select either [1-point Adjustment] or
[2 Point Adjustment].