Whenever you remove the balance enclosure,
make sure that the power is off and the
instrument is unplugged.
Figure 2.4
Removing the
Balance Enclosure
Only qualified service personnel should proceed
beyond this point.
10. Using a 7/64-inch hex wrench, loosen and
carefully remove the top and bottom clamp-
ing bars of the suspension shipping clamp.
11. Using the same wrench, remove the eight
(8) screws from the front and rear clamping
bars of the suspension shipping clamp.
Carefully slide the clamping bars horizontally
left or right to remove them.
12. Inspect the suspension components (wires
and springs) for damage.
13. Remove the shipping foam at the rear of the
balance assembly by pulling upward.