Using Your Options
TA I
NSTRUMENTS
TMA 2940 CE
4–22
To calculate the vial constant from a scan, use
the following equation:
β
ref
V
K =
(S B)
where:
K
= vial constant (cm
3
/µm)
S
= slope (
µ
m/
o
C)
β
ref
= cubic expansion coefficient of
standard material
m
3
m
3 o
C
β
ref
= 3
α
ref
Where
α
ref
is the coefficient of
linear expansion of standard
material, (m/m
o
C), measured
over the same temperature range.
V
= sample volume (cm
3
). This volume
may be obtained by dividing the
sample mass in grams by its density
in g/cm
3
at room temperature.
B
= slope of baseline (
µ
m/
o
C).
The dilatometer accessory kit contains:
Dilatometer probe assembly
Dilatometer sample vial
Filling medium
Aluminum calibration standards.
( )