RSA-G2 Getting Started Guide
Page 49
4 Left-to-right adjustment
: Viewing the level from the front of the instrument, extend the left-rear leveler
until the level’s air bubble aligns (left to right) with the circle on the glass face of the level.
Figure 32
Air bubble left-to-right aligned.
NOTE
: If left-to-right leveling is not possible due to the fixed right-rear foot on the instrument being
below level, it is necessary to either shim the table to level or shim the instrument’s fixed right back foot.
5 Front-to-back adjustment
: If necessary, extend both front levelers until they just touch the table top.
Figure 33
RSA-G2 front levelers.
Bubble aligned left to right
Front levelers