RSA-G2 Getting Started Guide
Page 22
NOTE
: Geometry baffles prevent heat from entering the transducer and motor areas. For optimum perfor
-
mance when running the FCO above 250°C, place the upper and lower geometry baffles outside the outer
top and bottom covers of the oven, as shown below. Refer to TRIOS Help for additional information.
Figure 10
RSA-G2 geometry baffles positioned properly above and below FCO, with sample loaded.
CAUTION: Make sure the purge air is flowing when the oven is on. Failure to do so could cause
damage to components inside the oven. The purge line connects to the back of the RSA-G2, to the
port labeled OVEN PURGE.
MISE EN GARDE: Assurez-vous que l'air de drainage circule lorsque le four est en marche. Le non
respect de cette précaution pourrait provoquer des dégâts sur les composants du four. La conduite
de drainage se raccorde à l'arrière de l'RSA-G2, au port étiqueté DRAINAGE DU FOUR.
Upper geometry baffle
positioned above FCO
Lower geometry baffle
positioned below FCO
Summary of Contents for RSA-G2
Page 1: ......
Page 9: ...Forced Convection Oven with tension clamp Technology 8...
Page 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...
Page 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...
Page 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...
Page 30: ...Verso Filler Page z z...
Page 31: ...n getting Started guide Revision D Issued July 2022...
Page 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...
Page 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...
Page 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...