RSA-G2 Getting Started Guide
Page 36
2
Connect the line from the filter to your air source.
3
Connect the outlet line from the regulator to the test station at the
Air
inlet port (see
location). This is an input line to the test station that is used to supply air to the oven, motor, transducer
air bearings, and throughout the test station.
4
Set the purge rate adjuster knob (shown below) to the setting number
2
.
Figure 18
Purge rate adjuster knob.
5
Turn on the air supply and use the pressure regulator knob to set the supply pressure to 80 to 90 psi
(0.55 to 0.62 MPa).
Figure 19