RSA-G2 Getting Started Guide 

Page 2

Notice

The material contained in this manual, and in the Help for the software used to support this instrument, is 

believed adequate for the intended use of the instrument. If the instrument or procedures are used for pur

-

poses other than those specified herein, confirmation of their suitability must be obtained from TA Instru

-

ments. Otherwise, TA Instruments does not guarantee any results and assumes no obligation or 

liability. TA Instruments also reserves the right to revise this document and to make changes without 

notice.
TA Instruments may have patents, patent applications, trademarks, copyrights, or other intellectual prop

-

erty covering subject matter in this document. Except as expressly provided in written license agreement 

from TA Instruments, the furnishing of this document does not give you any license to these patents, 

trademarks, copyrights, or other intellectual property.
TA Instruments Operating Software, as well as Module, Data Analysis, and Utility Software and their asso

-

ciated manuals and TRIOS Help, are proprietary and copyrighted by TA Instruments. Purchasers are 

granted a license to use these software programs on the module and controller with which they were pur

-

chased. These programs may not be duplicated by the purchaser without the prior written consent of TA 

Instruments. Each licensed program shall remain the exclusive property of TA Instruments, and no rights 

or licenses are granted to the purchaser other than as specified above.
TA Instruments can accept no liability for loss or damage, however caused, arising from the faulty or 

incorrect use of its products.TA Instruments shall not be liable for any damages caused by interactions 

between exogenous materials (e.g. chemicals) and parts of the instrument. This includes interactions of 

gaseous, liquid or solid materials with, for instance, ampoule surfaces and/or parts of the instrument. It also 

includes gases or vapors leaking from ampoules (e.g. originating from chemical reactions producing 

gaseous substances), with subsequent cause of damage to the calorimeter.

©2022 by 

TA Instruments — Waters LLC 

159 Lukens Drive

New Castle, DE 19720

Summary of Contents for RSA-G2

Page 1: ......

Page 2: ...ue multiwave arbitrary waveform and dielectric thermal analysis With such a broad range of solid analysis techniques the RSA G2 is uniquely positioned to address the widest range of applications from...

Page 3: ...ermine and ensure their suitability for applications processability and end use performance The RSA G2 is the ideal platform for fully characterizing and understanding complex mechanical behavior of s...

Page 4: ...se to external forces can be purely viscous or Newtonian behavior purely elastic or Hookean behavior or a combination of both Nearly all commercial materials of interest respond with a combination of...

Page 5: ...art of the RSA G2 dual head solids analyzer is the high performance drive motor and unique transducer The sample is deformed from the bottom by a direct drive DC Servo motor featuring all new electron...

Page 6: ...Transducer Motor Temperature Sensor Lower Geometry Mount Rare Earth Magnet __ Transducer Motor Air Bearings Air Bearings Drive Motor Upper Geometry Mount Technology 5...

Page 7: ...bearing eliminating backlash A linear optical encoder is mounted directly between the stationary frame and moving bracket for precision head positioning to an accuracy of 0 1 1m Benefits of the indepe...

Page 8: ...lysis during oscillation tests or post Fourier transformation analysis Higher odd harmonics that occur in the stress force signal in oscillation tests are a result of non linear response The ratio of...

Page 9: ...Forced Convection Oven with tension clamp Technology 8...

Page 10: ...n chamber to heat the sample quickly and without thermal gradients The FCO can be mounted on either side of the test station and comes standard with a long life intemal LED lamp and window viewing por...

Page 11: ...ding In this mode the sample is deformed around three point contacts at both ends and its middle It is considered a pure mode of deformation as the sample is freely supported by fulcrums eliminating c...

Page 12: ...e applied deformation is parallel to the sample thickness and the resultant deformation is simple shear Typical samples tested include polymer melts foams elastomers gels pastes and other soft solids...

Page 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...

Page 14: ...erature is observed to decrease by almost 29 C from 102 C to 73 C The figure on the right shows a series of frequency sweeps run on an Elastomer sample immersed in synthetic oil at a temperature of 25...

Page 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...

Page 16: ...ctric Bridges 25 mm Parallel Plates FCO Forced Convection Oven IEEE Internal to Instrument Keysight Model E4980AL 20 Hz to 1 MHz 0 001 to 2 Volts Keysight Model E4980A 20 Hz to 2 MHz 0 005 to 20 V The...

Page 17: ...s Viscoelastic materials exhibit a phase angle anywhere between these two ideal cases depending on the rate of deformation The figures below show these sinusoidal responses along with the variety of r...

Page 18: ...e strain amplitude At about 0 1 strain the modulus begins to decrease in magnitude showing the end of the LVR Outside the LVR the output stress response to the input strain is no longer sinusoidal Non...

Page 19: ...int or peaks in the E or tan b The temperature at which the transition occurs depends on the parameter selected 9 108 o CL Temperature Ramp on Polycarbonate tan b Peak Tg 156 8 C E Onset Tg 148 6 C E...

Page 20: ...re typically run over a limited range of 0 1 to 100 Hz as shown below on the right for a pressure sensitive adhesive at 70 C using the RSA G2 shear sandwich clamp At this temperature and frequency ran...

Page 21: ...emperature Superposition studies as all the frequency dependent data are collected at the same temperature Soak lime Oscillatory Measurement Time Time temperature superposition TIS is used to extend t...

Page 22: ...or fatigue studies The right figure shows an example of a two part epoxy cured on a supporting fiberglass braid using a dual cantilever clamp At short times the storage modulus is low and increases i...

Page 23: ...constant stress is applied to the sample at tl and the resulting strain is measured over time At t 2 the stress is removed and the recovery recoil strain is measured The creep compliance D t is calcu...

Page 24: ...ecovery was monitored for 60 s C C 0 0 Testing Modes and Applications 23 0 0 en l 5 0 0 2 c 0 0 X 0 a C Stress Relaxation on PDMS 106 105 104 103 102 Temperature 25 C 101 Strain 5 Shear Sandwich 10 10...

Page 25: ...eating rate followed immediately by an iso strain condition and linear cooling rate It can be observed in the figure that under iso stress the strain gets larger as the sample is heated and softens At...

Page 26: ...ow left figure shows a Polyethylene film tested using the tension geometry at 25 C and a constant linear rate of 5 I Im s A classical stress strain curve is observed with yield point ultimate tensile...

Page 27: ...tures as low as 55 C and 100 C respectively Utilizing compressed air the Air Chiller Systems can help eliminate or reduce liquid nitrogen usage from any laboratory and offers an incredible return on i...

Page 28: ...Temperature Range Heating Rate Cooling Rate Isothermal Stability SPECIFICATIONS 27 SPECIFICATION S I RSA G2 0 0005 N 35 N 0 00001 N 0 00005 to 1 5 mm 1 Nanometer 103 to 3 X 1012 P 1 0 0001 0 00001 2 x...

Page 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...

Page 30: ...Verso Filler Page z z...

Page 31: ...n getting Started guide Revision D Issued July 2022...

Page 32: ...ata Analysis and Utility Software and their asso ciated manuals and TRIOS Help are proprietary and copyrighted by TA Instruments Purchasers are granted a license to use these software programs on the...

Page 33: ...ual Supplement Please click the TA Manual Supplement link to access the following important information supplemental to this Getting Started Guide TA Instruments Trademarks TA Instruments Patents Othe...

Page 34: ...2 010 Particular requirements for laboratory equipment for the heating of materials For European Economic Area In accordance with Council Directive 2006 95 EC of 12 December 2006 on the harmonization...

Page 35: ...ntrol and laboratory use EMC requirements Part 1 General Requirements Emissions Meets Class A requirements per CISPR 11 Immunity Per Table 1 Basic immunity test requirements For the United States CFR...

Page 36: ...VERTISSEMENT L utilisateur de cet instrument est pr venu qu en cas d utilisation contraire aux indications du manuel la protection offerte par l quipement peut tre alt r e Symbol Explanation This symb...

Page 37: ...l parts must be performed only by TA Instruments or other qualified service personnel Ce symbole appos l arri re du panneau d acc s indique vous devez d brancher l instrument avant d effectuer des tra...

Page 38: ...vous n tes pas form aux proc dures lectriques ne d posez pas les couvercles de l armoire sauf indications sp cifiques contenues dans le manuel La maintenance et la r paration des pi ces internes doive...

Page 39: ...9 Mechanical Safety DANGER Keep your hands clear of the motor while it is running If you need to touch it turn OFF the motor first DANGER loignez vos mains du moteur lorsqu il tourne Si vous devez le...

Page 40: ...Overview 13 System Components 13 RSA G2 Test Station 14 Stage 14 RSA G2 Keypad 14 RSA G2 Touch Screen 16 The RSA G2 Instrument Touch Screen 16 The RSA G2 Status Touch Screen 17 The RSA G2 Messages Tou...

Page 41: ...sure Specifications 38 Air Quality 38 Making the Test Station Purge Connections 38 Connecting the Signal Panel 38 Connecting Cables and Lines 41 Connecting the Modular Power Supplies 41 Test Station L...

Page 42: ...uctions for Geometry Installation 60 General Guidelines 60 Installing the Upper Geometry 61 Installing the Lower Geometry 62 Creating a Procedure 63 Running an Experiment 64 Maintaining the Instrument...

Page 43: ...ller TA Instruments TRIOS software and a Forced Con vection Oven FCO to make up a rheological system Note that the FCO is the only environmental system that can be used with the RSA G2 Your controller...

Page 44: ...t station the touch screen or through TRIOS software control The rate at which the stage moves can be adjusted through software RSA G2 Keypad The keypad shown in the figure below is an external device...

Page 45: ...on tells you the state of the motor If the transducer and motor are both off lights are red pressing the motor button will also reset the transducer if a force overload occurred and caused it to turn...

Page 46: ...the interactive functions shown on the touch screen displays NOTE For detailed information on the functions performed with the touch screen not found in this manual see TRIOS software online help The...

Page 47: ...vides easy access to system information and diagnostics that may be used to troubleshoot your instrument Figure 4 RSA G2 Status touch screen The primary use of this screen is to provide an overview of...

Page 48: ...location and network IP address infor mation In addition you can adjust FCO camera focus and light and calibrate the touch screen Figure 6 RSA G2 Settings touch screen Linear Transducer The RSA G2 ut...

Page 49: ...d scaled to provide a force value Selection and control of the force range is performed using the TRIOS software Note that the maximum force measurable by the transducer is 3500 grams For axial force...

Page 50: ...station needs to be moved or in the event that the air supply will be removed for an extended period of time Note that the motor cover must be removed to expose the threaded hole that is used to inst...

Page 51: ...ouch the outer cover when the oven is running as the outer covers can get very hot Only touch the FCO handle if necessary When opening the oven to load samples the heaters and the gas flowing through...

Page 52: ...below FCO with sample loaded CAUTION Make sure the purge air is flowing when the oven is on Failure to do so could cause damage to components inside the oven The purge line connects to the back of th...

Page 53: ...Temperature testing above 500 C is limited to twohours Transducer may shut down ifrun for a longer time period The FCO has two forced gas gun heaters with counter rotating air flow for a wide tempera...

Page 54: ...RT used for fast response temperature control Using the default system settings the RSA G2 reports and controls the measured temperatures using the mid oven air PRT temperature Measured oven temperatu...

Page 55: ...k a solenoid valve to control liquid nitrogen flow into the Dewar and hardware to control the boiling of the liquid nitro gen to produce cold nitrogen gas within the flask A second solenoid valve cont...

Page 56: ...rnace 100 C Discovery Hybrid Rheometer models with Environmental Test Chamber 85 C and ARES G2 RSA G2 with Forced Convection Oven 100 C The ACS 3 features low noise durable compressors approx 55 dB sm...

Page 57: ...before being mounted on Air Chiller Table 3 ACS 3 and Chiller Panel Specifications Dimensions of Air Chiller H x W x D Without Chiller Panel With Chiller Panel 112 cm 44 in x 37 cm 14 5 in x 56 cm 22...

Page 58: ...us que le tuyau d arriv e d air et tous les raccordements lectriques sont correctement effectu s NOTE For Camera accessory installation instructions refer to TRIOS software Help The FCO can be optiona...

Page 59: ...er Additionally factors such as anisotropy and differences in strain dependence may yield inconsistent results for different geometries Available Geometries The following is a list of the geometries t...

Page 60: ...sure 104 5 kg 230 lbs 12 7 kg 28 lbs 6 8 kg 15 lbs Power Test station Power supply enclosure Oven supply enclosure From power supply 90 264 VAC 16 amps max 180 264 VAC 47 63 Hz 13 amp max Operating en...

Page 61: ...parts Choosing a suitable location Connecting the RSA G2 to a suitable air supply Connecting all cables and lines Connecting the RSA G2 to the customer line voltage outlet Connecting the RSA G2 to the...

Page 62: ...e power and all power cords and cables to the power supply c Lock the motor and transducer d Turn off the air gas supply and disconnect all lines 2 Remove the four corner plugs by pulling them straigh...

Page 63: ...G2 without a cart you must have three people to perform the following procedure 1 The handling points on the instrument are shown Grasp the instrument at those points Figure 16 Instrument handling po...

Page 64: ...e open and closed positions of all heated environment control systems An area that does not restrict airflow to cooling fans on the rear of the power control box or the option controllers On A stable...

Page 65: ...air should be 10 C or lower We highly recommend installing the optional air dryer filter between the test station and the air supply If your air supply has excessive moisture levels that result in imm...

Page 66: ...r its location This is an input line to the test station that is used to supply air to the oven motor transducer air bearings and throughout the test station 4 Set the purge rate adjuster knob shown b...

Page 67: ...roller AIR inlet Connected to the external air dryer This port is used to supply dry air to the oven motor transducer air bearings and throughout the test station The air supply pressure should be 80...

Page 68: ...orm your laboratory manager or compressor maintenance personnel of your instrument requirements in detail You should also ask to be informed before any air supply interruption or compressor maintenanc...

Page 69: ...pled and stored in the experiment data file This input signal is defined using TRIOS software ANALOG 1 OUT A 10 VDC signal that is defined using TRIOS software ANALOG 2 OUT A 10 VDC signal that is def...

Page 70: ...channel instead of the motor strain signal User selectable Input scaling is 5 VDC NORMAL IN An external user input that can be connected to the internal normal measure ment channel instead of the tran...

Page 71: ...is serrage main sur tous les c bles d ordinateur Chaque fois que vous bran chez ou d branchez les cordons d alimentation tenez les par les fiches et non par les cordons Connecting the Modular Power Su...

Page 72: ...nected 3 Connect one end of the largemulti pin cable to the test station connector The other end will be connected to the RSA G2 test station later in this manual 4 Plug in up to four accessory power...

Page 73: ...supply Figure 24 FCO controller power supply 3 Oven Connect the power cable from the oven to the connector to the left of the fan 4 Obtain the loose end of one of the power cables that are already con...

Page 74: ...lk tank Bulk tank needs to be a low pressure 22 psi source 2 Obtain the LN2 transfer line 3 Connect the elbowed end of the transfer line to the LN2 Controller 4 Connect the opposite end of the transfe...

Page 75: ...at the software knob is set for DHCP 2 Using keyboard icons enter the Machine ID and Location 3 Power down the RSA G2 4 Make sure the host computer LAN is set for DHCP 5 Ensure that TRIOS Instrument C...

Page 76: ...mputer to LAN If you wish to also connect the computer to a LAN you will need to have already installed a network interface card 1 Locate the second Ethernet port on the back of the computer 2 Plug on...

Page 77: ...n Once all cables and lines have been connected to the modular power supplies and the RSA G2 test station you can connect the power cable Install the power cable as follows 1 Make sure the power switc...

Page 78: ...its location Counterclockwise rotation of the leveler retracts it lowering the instrument at its location Figure 30 RSA G2 leveler Follow the instructions below to perform this procedure 1 Make sure...

Page 79: ...Figure 32 Air bubble left to right aligned NOTE If left to right leveling is not possible due to the fixed right rear foot on the instrument being below level it is necessary to either shim the table...

Page 80: ...nd both front levelers equally until the level s air bubble aligns front to back with the circle on the glass face of the level Figure 34 Air bubble front to back aligned NOTE By extending both levele...

Page 81: ...ion After this is done the LED near the Standby button on the keypad should be solid red 3 Press the standby button on thetest station front panel shown in the figure below The instrument power status...

Page 82: ...instrument is expected to be disassembled for maintenance disconnected from its compressed air source or is being moved then it is necessary to turn the power to the system completely off Please foll...

Page 83: ...priate geometry for the desired test and sample Creating or choosing the test procedure and entering sample and instrument information through the TRIOS Instrument Control Software Loading the prepare...

Page 84: ...details on how to perform that calibration refer to the online documentation accessed through the TRIOS software The table below lists required tasks and the recommended interval for each task Table 7...

Page 85: ...TRIOS software Refer to TRIOS Help for additional information Table 8 RSA G2 Geometry Mass Size mm Type Position Mass mg Compliance mrad N m 3 pt bending top 69 0 10 3 pt bending bottom 84 0 15 3 pt...

Page 86: ...e of steel is near zero If the values obtained from this test are abnormal please contact Technical Services for further assistance Force Calibration Perform a Force Calibration using TRIOS software R...

Page 87: ...e best tested using the clamped bending geometry Above the glass transition these samples can be tested in either compres sion using the compression parallel plate geometry or in shear using the shear...

Page 88: ...than 10 percent of the com manded strain If actual strain is much lower the sample needs to be made less stiff by lengthening it and making it narrower and or thinner If necessary change the geometry...

Page 89: ...onofilament Linear tension cylindrical Length up to 55 mm1 optimal length 35 40 mm Diameter up to 0 5 mm Film Linear tension rectangular Length up to 55 mm1 optimal length 35 40 mm Width up to 12 7 mm...

Page 90: ...e sample Make sure that the sample PRT does not actually touch the sample or geometry Close the oven care fully until you are sure the sample PRT is adjusted properly The alignment pins allow the geom...

Page 91: ...nd concentric alignment Holding the geometry by the shaft align the notch on the geometry with the slot on the transducer mount and insert the geometry into the transducer mount Refer to Figure 36 for...

Page 92: ...or dust at the location surfaces for parallel and concentric alignment Holding the geometry by the shaft align the notch on the geometry with the slot on the motor mount and insert the geometry into...

Page 93: ...al functions that are obtained are dependent on frequency Step Transient A step test increments to a target value of stress creep strain stress relaxation or strain rate and measures the response as a...

Page 94: ...other than ambient install the geometry and then adjust the temperature to the desired initial value Allow the geometry to reach thermal equilibrium 2 Enter the desired sample information using TRIOS...

Page 95: ...r power supply module from the wall outlet 3 Inspect all the cords for frayed insulation or exposed bare copper wire especially in the immediate vicinity of the plugs on either end If any damage is fo...

Page 96: ...the air dryer can be checked and changed Make sure the moisture indicator is green If it turns red disconnect the air line drain the filter bowls and change the filter cartridges Make sure the air dr...

Page 97: ...e lower power supply module to the On position Cleaning the Touch Screen Clean the touch screen as often as desired Clean the touch screen with a household liquid glass cleaner and soft cloth Wet the...

Page 98: ...s for the RSA G2 are listed below Refer to the table below when ordering parts Table 10 RSA G2 Replacement Parts Part Number Description Replacement coalescing filter element 613 06404 Replacement par...

Page 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...

Page 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...

Page 101: ...t OEM Customer License Agreement for Embedded Systems 6 Attachment B to End User License Agreement TRIOS Optional Software 9 Attachment C to End User License Agreement Discovery Thermal Instruments 3r...

Page 102: ...ftware as a Demo or Evaluation version then your license rights under this EULA shall extend solely for the evaluation period You acknowledge that any Demo or Evaluation version may not be at the same...

Page 103: ...ntal special or consequential damages loss of data or data being rendered inaccurate loss of profits or revenue or interruption of business in any way arising out of or related to the use or inability...

Page 104: ...esses for which the DEVICE has been designed and marketed by TA Instruments This license specifically prohibits any other use of the software programs or functions or inclusion of additional software...

Page 105: ...upport services related to the SOFTWARE MS Microsoft Corporation and their affiliates may use this information solely to improve their products or to provide customized services or technologies to you...

Page 106: ...nstruments on separate media and labeled Recovery Media you may use the Recovery Media solely to restore or reinstall the SOFTWARE originally installed on the DEVICE Backup Copy You may make one 1 bac...

Page 107: ...on Linear Analysis Package This data analysis option may be installed on the primary instrument control computer as well as one other computer provided the software is only used for analysis of data g...

Page 108: ...taining the applicable 3rd party license agreement Note that these PDF documents are available only through the installation of TRIOS Instrument Control Software and its associated documentation Adobe...

Page 109: ...ITCRun is a trademark of TA Instruments Waters LLC 159 Lukens Drive New Castle DE 19720 BindWorks is a trademark of TA Instruments Waters LLC 159 Lukens Drive New Castle DE 19720 TA Instruments Patent...

Page 110: ...ibes proprietary technology patented by TA Instruments Waters LLC U S Patent No 6 931 915 Platen for Automatic Sampler describes proprietary technology patented by TA Instruments Waters LLC U S Patent...

Page 111: ...C U S Patent No 5 973 299 Metallic Sheet Insulation System describes proprietary technology patented by TA Instruments Waters LLC U S Patent No 5 915 283 Dynamic and Thermal Mechanical Analyzer Having...

Page 112: ...nformation on our latest products contact information and more see our Web site at http www tainstruments com TA Instruments Waters LLC Corporate Headquarters 159 Lukens Drive New Castle DE 19720 USA...

Page 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...

Page 114: ...tents 2 Ideal Setup 3 System Components 4 Instrument Measurements 5 Utility Requirements 6 Computer Requirements 7 8 Hardware 7 Software 8 Accessories 9 ACS and Chiller Panel 9 11 Air Dryer 12 Liquid...

Page 115: ...ation with adequate floor space and a rigid laboratory bench that is level and is in a vibration free environment For optimal performance it is recommended that the instrument be placed by itself on a...

Page 116: ...Revision G Issued May 2022 Page 4 MAIN SYSTEM COMPONENTS A B C A Computer Controller B Instrument C Power Supply Enclosure D Forced Convection Oven Enclosure E1 Liquid Nitrogen Controller LN2 Tank OR...

Page 117: ...MAIN INSTRUMENT POWER SUPPLY ENCLOSURE FORCED CONVECTION OVEN ENCLOSURE Height 104 cm 39 37 in Height 34 cm 13 5 in Width 32 cm 12 5 in Depth 48 cm 19 in Depth 56 cm 22 in Width 46 cm 18 in Weight 87...

Page 118: ...be free from oil 0 01 mg m3 and dirt 5 m Dew Point 10 C or better Pressure 100 psig 0 7 MPa Flow Rate 9 5 scfm 270 L min If using N2 gas as FCO heater source air flow rate is 6 scfm 170 L min Separate...

Page 119: ...IOS Screen resolution Required 1280 x 1024 with 24 bit colors Recommended 1920 x 1080 with 24 bit colors Graphic memory 128 MB Screen LCD size Required 19 or greater Recommended 24 wide screen USB II...

Page 120: ...Pack Microsoft Operating System Service Pack Updates Windows Operating System and associated Microsoft updates must be up to date Windows 10 must be 1709 or later Network A second network card for co...

Page 121: ...MENTS Height 88 5 cm 35 in Depth 56 cm 22 in Weight 96 kg 211 lbs WITHOUT Chiller Panel Width 67 cm 26 5 in Weight 112 kg 247 lbs WITH Chiller Panel Height 112 cm 44 in Width 67 cm 26 5 in Depth 56 cm...

Page 122: ...FCO CHILLER PANEL MEASUREMENTS SMC MODEL FCO CHILLER PANEL MEASUREMENTS PARKER MODEL DISCONTINUED Height 68 6 cm 27 in Width 54 6 cm 21 5 in Depth 20 3 cm 8 in Weight 8 2 kg 18 lbs Height 86 4 cm 34 i...

Page 123: ...erial number plate on the rear of the unit US sites require an L6 20 single phase outlet Gas Air or nitrogen Pressure 6 9 bar 100 psig Flow rate 200 L min Temperature 20 30 C Dew point Must not exceed...

Page 124: ...bs It has two mounting holes 22 3 cm 8 8 in apart Must be mounted upright to the wall within 183 244 cm 6 8 ft of the air source Customer must provide Means to connect to a 3 8 NPT male connector on t...

Page 125: ...psig 207 kPa gauge may cause damage to the Liquid Nitrogen Controller Keep the supply line short and provide adequate insulation to minimize gaseous nitrogen build up in the supply line Failure to do...

Page 126: ...personnel will be on site the day of installation Air Supply Pressure is 100 psig 0 7 MPa Available flow rate is 9 5 scfm 270 L min Dew point is 10 C or better Liquid Nitrogen Controller Customer mus...

Page 127: ...and more see our website at http www tainstruments com To find your local TA Instruments office and contact information visit http www tainstruments com contact ta directory TA Instruments Waters LLC...