LIQUID NITROGEN CONTROLLER MEASUREMENTS
LIQUID NITROGEN CONTROLLER REQUIREMENTS
Requirements
Should be placed on the same side as the FCO
Customer must provide: 160 L (or larger) Liquid Nitrogen tank with a pressure of
14–22 psig (95–150 kPa gauge)
LN
2
pressure above 30 psig (207 kPa gauge) may cause damage to the Liquid
Nitrogen Controller.
Keep the supply line short and provide adequate insulation to minimize
gaseous nitrogen build-up in the supply line. Failure to do so may cause the
Liquid Nitrogen Controller to malfunction frequently. Use the 3 ft. hose
provided by TA Instruments for this reason.
Circulator Power Cooling
Gas
LN
2
Fluid
Light Hardware Software Temp
Lab
Customer
Weight when EMPTY: 14 kg (30 lbs)
Height: 58 cm (23 in)
Width: 28 cm (11 in)
Depth: 36 cm (14 in)
Weight when FULL: 15 kg (33.5 lbs)