10. When you are satisfied that the pan is
properly positioned, press the
CALIBRATE
key repeatedly until you see the correct cell
pan position on the display:
Calibrate/Verify
Cell Ref. Pos.
11. Press
ENTER
. The screen displays:
Verify (1) or
Store (2) ? 1
12. Select option 2 (Store) to store the calibra-
tion position (press the 2 key, then press
ENTER
). The screen displays:
Cell Ref. Pos.
Calibrated
13. Release the pan (press the
RELEASE
key).
Press
RESET SAMPLER
and
ENTER
to
bring the Autosampler CE to the home
position. Press the
CALIBRATE
key until
you see the correct cell lid position on the
display:
Calibrate/Verify
Cell Ref. Pos.
14. Press
ENTER
. The screen displays:
Verify (1) or
Store (2) ? 1
15. Press
ENTER
to select default option 1
(Verify).
Performing the Calibration
Summary of Contents for DSC Autosampler CE
Page 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Page 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Page 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Page 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...