Diagnostics
6–34
TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Error
Pan #xx dropped
Problem: The Autosampler CE has dropped a
pan or cannot pick it up (xx indicates
the sample tray number). This can be
caused by (1) incorrect calibration, (2)
bent or damaged gripper fingers, or (3)
a faulty Autosampler CE gripper
mechanism.
Solution: Try recalibrating the related
position(s); then repeat the task. If the
error occurs again, call a qualified
service representative.
Error
Pan #xx missing
Problem: A pan that the Autosampler CE is
attempting to load or unload is missing
from the DSC cell or the tray.
Solution: Whether the pan is missing from the
cell or the tray, return it to its proper
sample tray position, and return all
other pans in the cell to their proper
tray positions. Press
Reset
, then
select
Start AS-Run.
Enter the run
number of the failed run to retry the
task.
Summary of Contents for DSC Autosampler CE
Page 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Page 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Page 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Page 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...