Diagnostics
6–18
TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
The GPIB Test
The GPIB test enables you to verify the GPIB
address setting for the Autosampler CE (for
further information, see Chapter 2).
1. Access the Autosampler CE keypad by
loosening the captive thumbscrews on the
keypad cover and tilting the cover down.
2. Press the
TEST
key six times. The Auto-
sampler CE screen displays:
Test
GPIB
3. Press
ENTER
to select the test displayed on
the screen. The screen should display a
numeral ten higher than the address you set
when you configured your DSC:
GPIB
Address = xx
For example:
DSC 2920 address = 1
Autosampler CE address = 11
4. Press
ESCAPE/CLEAR
to exit the GPIB
test.
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.
NOTE:
Summary of Contents for DSC Autosampler CE
Page 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Page 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Page 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Page 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...