4–33
Step B:
Use the LIDS key to test the Autosampler
CEs ability to cover the DSC cell.
1. Press
RESET SAMPLER
and
ENTER
to
bring the Autosampler CE to home position.
2. Press the LIDS key. The Autosampler CE
screen displays:
Uncover (1) or
Cover (2) ? 1
3. Select option 2 (Cover) by pressing the 2
key, then pressing
ENTER
. The Autosam-
pler CE removes all lids from their storage
positions and places them on the cell.
4. Press the
LIDS
key again. The screen
displays:
Uncover (1) or
Cover (2) ? 1
5. Select default option 1 (Uncover) by press-
ing
ENTER
. The Autosampler CE removes
all lids from the cell and returns them to their
storage positions.
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.
Checking Your Calibration
NOTE:
Summary of Contents for DSC Autosampler CE
Page 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Page 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Page 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Page 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...