6–33
Solution: Return the lid to its storage position.
When the Autosampler CE senses the
lid, it should return to normal operation.
If this does not work, test the lid
storage sensor with the
TEST
key
(see page 6-9).
If this error message appears when
the Autosampler CE is gripping a lid,
the gripper sensor may be malfunc-
tioning; test the gripper sensor with the
TEST
key.
Error
Lid #x not strd.
(Lid #x not stored)
Problem: The Autosampler CE has been reset
with one of the Reset Sampler keys,
and one of the lids (lid #1, #2, or #3)
has not been returned to its storage
position.
Solution: Use tweezers to return the lid to its
storage position. If the Autosampler
CE is in manual mode, press
ES-
CAPE/CLEAR
to clear the error
message from the screen.
Error Messages
Summary of Contents for DSC Autosampler CE
Page 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Page 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Page 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Page 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...