Performing the Calibration
Calibrating the
Cell Lid Positions
Because each DSC cell is custom made, the cell
lid positions cannot be calibrated at the factory
and must be calibrated on newly installed
Autosampler CEs.
1. Access the Autosampler CE keypad by
loosening the captive thumbscrews on the
keypad cover and tilting the cover down.
2. Press
RESET SAMPLER,
and press
ENTER
to bring the Autosampler CE to
home position.
3. Make sure all lids are in their storage
positions (see Table 4.1).
4. Press the
LIDS
key. The screen displays:
Uncover (1) or
Cover (2) ? 1
5. Select option 2 (Cover) by pressing the 2
key, then pressing
ENTER
. The Autosam-
pler CE picks up a lid and begins moving
toward the cell.
6. Press the
ESCAPE/CLEAR
key when the
Autosampler CE reaches the DSC cell area.
7. Use the motion keys to position the lid in its
correct cell calibration position (see Table
4.1). Press the
RELEASE
and
GRIP
keys
repeatedly to ensure that the fingers can
firmly grasp the lid, and watch carefully to
ensure that each finger contacts the lid at the
same time (see Figure 4.1).
Summary of Contents for DSC Autosampler CE
Page 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Page 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Page 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Page 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Page 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...