technical reference
3
29
Synrad
Firestar i401
operator’s manual
Integrating Firestar safety features
Your control system can monitor the laser’s shutter status on the
User I/O
connector by connecting your
system’s input to Pin 14,
Shutter Open
, and Pin 13,
Output Common
(see Figure 3-16). The
Shutter
Open
output closes when a
Shutter Open Request
signal is present (
SHT
LED illuminated blue) and the
Laser Ready
output is closed (
RDY
LED is On). The output is open (
SHT
LED Off) when the
Shutter
Open Request
signal is removed or the
Laser Ready
output is open (
RDY
LED is Off).
Remote interlock functions
Interlock circuits are often used to disable machinery when a shield, panel, or door is opened. Firestar’s re-
mote interlock function allows you to connect into an external remote interlock circuit and prevent lasing
by removing DC power from the laser’s RF driver boards when the circuit is electrically “open”.
Lasing is enabled when a
Remote Interlock
signal is present (
INT
LED illuminated green), if the
RDY
LED is illuminated and a
Shutter Open Request
signal is applied. Lasing is disabled when the
Remote
Interlock
signal is removed (
INT
LED red,
RDY
LED off). DC power is applied to the RF driver only when
the
INT
LED is green and the
RDY
LED is yellow. Remote interlock functionality is provided by the
Re-
mote Interlock
signal via Pin 3 on the
User I/O
connector.
To use Firestar’s remote interlock feature to initiate lasing, apply a voltage in the range of ±5–24 VDC to
Pin 3,
Remote Interlock
. Applying a
Remote Interlock
signal causes the
INT
LED to turn green, the
RDY
indicator to turn yellow, and sends DC power to the laser’s RF boards. After a five-second delay, a tickle
signal is applied to the tube. When a
Shutter Open Request
signal is present, PWM Command signals
are enabled to begin lasing. Removing voltage stops DC power from reaching the RF driver, causing the
INT
LED to turn red and the
RDY
LED to turn Off. Lasing remains disabled until a voltage is reapplied to
Pin 3.
Your control system can monitor the laser’s remote interlock status on the
User I/O
connector by connect-
ing your system’s input to Pin 15,
Interlock Open
, and Pin 13,
Output Common
(see Figure 3-16). This
output is closed when remote interlock circuitry is open (
INT
LED illuminated red). The output is open
(
INT
LED green) when interlock circuitry is closed.