
Parameter setting
Page 52/55
SYCLOPE TERE’O Touch XL
®
Installation, starting, and programming instructions
a)
pH
Pressing the « Calibration » button opens the programming window.
Offset
Gain
Erasing
Calibration principle is identical for pH measurement or chlorine measurement.
-
To calibrate the pH in 1 point on a photometer reading, use the
Offset
tab
.
-
If the calibration is done at pH7 we will use the
Offset
tab
.
-
If the calibration is done at pH4 or 9 we will use the
Gain
tab
.
-
To calibrate the chlorine in 1 point on a reading photometer we will use the
Gain
tab
.
-
Offset tab of chlorine is very rarely used unless requested by an experienced technician.
-
To return to the "factory" version, use the
Clear
tab
.
Offset or pH7 calibration
Select the Offset tab.
Press the key to open the input keyboard of the new near pH 7.
Press calibrate to validate the entry.
-
In case of correct calibration "Calibration OK" is displayed in green
-
In case of calibration error "Error Calibration" is displayed in red
Gain or pH4 calibration
Select the Gain tab.
Press the key to open the input keyboard of the new near pH 4.
Settings
pH channel
Calibration
Calibration type selection
Measured value
7.18
Enter the standard value
7.18
Calibrate
Offset Gain Clear
Settings
pH channel
Calibration
Calibration type selection
Measured value
9.20
Enter the standard value
9.20
Calibrate
Offset Gain Clear
Settings
pH channel
Calibration
Calibration type selection
Factory calibration
Will be restored
Clear
Offset Gain Clear
Offset
Calibrate
Gain