SYCLOPE electronic TERE'O Touch XL Installation And Operating Instructions Manual Download Page 45

 Parameter setting 

 

Page 45/55 

 

SYCLOPE TERE’O Touch XL

® 

Installation, starting, and programming instructions

 

 

 

Signal level messages 

 

Searching …:

 Waiting for the Modem module answer. 

 

Error:

 No answer from the modem, check the hardware connection of the module. 

 

Low:

 No sufficient level to make the connection. 

 

Medium, Good, Excellent:

 Correct answer expected by the system. 

 

 

Ethernet Modem: 

 

State messages 

 

Searching…:

 Waiting for the Modem module answer. 

 

Error:

 No answer from the modem, check the hardware connection of the 

module. 

 

Waiting:

 Waiting for network connection. 

 

Negotiation:

 Connection in progress. 

 

Authentication:

 Connection in progress. 

 

Disconnecting:

 Disconnecting in progress. 

 

Disconnected:

 System not connected, check your TCPIP configuration 

 

Connected:

 Correct answer expected by the system. 

 

 

IP messages 

 

Searching…:

 Waiting for the Modem module answer. 

 

Error:

 No answer from the modem, check the hardware connection of the 

module. 

 

xxx.xxx.xxx.xxx: 

Current IP of your controller, a value other than 0.0.0.0 

is correct. 

 

 

WIFI Modem: 

 

State messages 

 

Searching…:

 Waiting for the Modem module answer. 

 

Error:

 No answer from the modem, check the hardware connection of the 

module. 

 

Waiting:

 Waiting for network connection. 

 

Negotiation:

 Connection in progress. 

 

Authentication:

 Connection in progress. 

 

Disconnecting:

 Disconnecting in progress. 

 

Disconnected:

  System  not  connected,  check  your  TCPIP  &  WIFI 

configuration 

 

Connected:

 Correct answer expected by the system. 

 

 

IP messages 

 

Searching…:

 Waiting for the Modem module answer. 

 

Error:

 No answer from the modem, check the hardware connection of the 

module. 

 

xxx.xxx.xxx.xxx: 

Current IP of your controller, a value other than 0.0.0.0 

is correct. 

 

 

Signal level messages 

 

Searching…:

 Waiting for the Modem module answer. 

 

Error:

 No answer from the modem, check the hardware connection of the 

module. 

 

Low:

 No sufficient level to make the connection. 

 

Medium, Good, Excellent:

 Correct answer expected by the system. 

 

 
 

j)

 

Initialization 

 

Summary of Contents for TERE'O Touch XL

Page 1: ...SYCLOPE TERE O Touch XL Controller for public swimming pool Part 1 Installation starting and programming instructions Reference DOC0464 Rev 2 0...

Page 2: ...rmation SYCLOPE Electronique 2019 2020 Manual of 27 03 2020 Rev 2 0 Analysers Controllers for swimming pools Product line TERE O Touch Installation starting and programming instructions Editor SYCLOPE...

Page 3: ...13 1 Installation conditions 13 2 Wall installation of the device 13 3 Open Close transparent door 14 4 Open Close the terminal cover 14 5 Electrical connection 15 6 Changing the internal fuse of P3...

Page 4: ...chlorine or bromine Thanks to HYDRO TOUCH ease of use their user friendliness and their remarkable technicality you will fully enjoy its many possibilities and will be assured of a perfect control and...

Page 5: ...the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful...

Page 6: ...ructions Information Action to do Element of a list or enumeration 4 Symbols and signs Identification of a continuous voltage or current Identification of an alternative voltage or current Protective...

Page 7: ...able glands conform to the maintenance of IP65 protection Cables used must be adapted to them in order to respect the protection index Shielded cables for connecting pH and ORP electrodes are not supp...

Page 8: ...work safety and prevention of accidents Are trained in the use of the device and its environment Have read and understood these instructions warnings and manipulation rules 3 Risk prevention The inst...

Page 9: ...tification and localization of the nameplate Manufacturer s label Particular risk Read the manual Model of the product Product which can be recycled Reference of the product Limitation of dangerous su...

Page 10: ...ions consumers within the European Union are required as of this date to return their used devices to the manufacturer who will take care of disposing them at no extra expense According to European di...

Page 11: ...3 2 Harmonics tests comply with EN61000 3 3 Immunity tests comply with EN61000 4 2 Immunity tests comply with EN61000 4 3 Immunity test EN61000 4 4 Immunity tests comply with EN61000 4 5 Immunity tes...

Page 12: ...ms Expressed in actual measurement value High and low threshold control Controlling Flowrate control Controlling injection to control water circulation Tank level Controlling injection to control the...

Page 13: ...ration on a suitable support and with no deformation If these instructions are not observed The controller risks to be damaged The measurements can be disrupted The warranty is not applicable 2 Wall i...

Page 14: ...tly To open the transparent door Door locked Lift the lock and pull towards the front of the device Door opened To close and lock the transparent door Put your fingers behind the lock and bring the do...

Page 15: ...act with neighbouring cables Secure the wire connections on the terminal blocks using a cable tie SYCLOPE TERE O Touch must be slaved to the filtration of the swimming pool using digital input Interna...

Page 16: ...n flat cable connecting the bottom card and the upper part of the device Locate the fuse to be changed on the lower PCB Remove the protective cap Change the fuse and replace the protective cap Reconne...

Page 17: ...he earth on the PL1 stud using an M4 eyelet terminal Tighten the cable gland to seal Your SYCLOPE TERE O Touch doesn t have a power switch So it s directly supplied when it s connected to the mains 8...

Page 18: ...n N 44 of the P4 mains terminal block Connect the earth on PE 45 of the P4 mains terminal block Tighten the cable gland to seal 9 Potential free relay connections P2 The potential free relay outputs c...

Page 19: ...as regulation for pH Use a 2 wire cable with a section appropriate for the voltage and current Remove the protective sheath Strip wires on 7mm Pass the cable through a cable gland Connect a cable on...

Page 20: ...r bromine measurement a pH input Use a coaxial shielded cable supplied with your sensor Connect the core of the cable to the Pot 1 connection Connect the shield on the Ref 2 connection Tighten the cab...

Page 21: ...8 Wire the other strand on Iin2 9 Tighten the cable gland to seal NOTE Be careful to respect the polarity of the sensor d Power input When necessary it s possible to use chlorine or bromine sensor re...

Page 22: ...ng contact system controlled by the filter motor or a flow sensor the following electrical diagram must be made Preferably use a two strand cable Wire a strand on Aux sw 4 Wire the other strand on Aux...

Page 23: ...n at the bottom of the tank bottom input so as to cut the injection when the tank is empty Your SYCLOPE TERE O Touch has a software configuration of the direction of contact You can use either a NO or...

Page 24: ...n SYCLOPE TERE O Touch has an RS485 communication port for linking a desktop computer equipped with a 485 port and a communication software to record measurement value alarms and different states of t...

Page 25: ...sult the instructions of this converter for the connection Reference Name INF1021 USB 485 Converter The controllers can be chained by respecting the order of the cables putting in parallel b Polarizat...

Page 26: ...is recommended in the event of single swimming pool where filtration circuit is independent Water is sampled after the recirculating pumps and before the filter entry If using before flocculent injec...

Page 27: ...regulation processes are inactive SYCLOPE TERE O Touch controller is delivered with standard programming It is advisable for the user to modify this programming if it does not correspond to the needs...

Page 28: ...modem installed On Off control button System in operation System stopped Pressing on this area of the screen stop or start your controller b Display of a measurement channel Time Type of measure Progr...

Page 29: ...ountdown Off No timer or delay completed NOTE The fixed ON state is only available from software version 3 04 Warning direct calibration is only possible if the system is in START mode If the User Pas...

Page 30: ...a central indication area of the selected element b Mode numeric When a programming element is numeric a numeric keypad appears when typing In the case of an input error or out of range value the Min...

Page 31: ...character key pressed to bring up the list of characters move your finger from left to right and release on the desired character Numeric keypad Special characters keyboard Press on to toggle the disp...

Page 32: ...ccess the INSTALLER menu When prompted for a password you can enter 0000 to access the service window This is valid from version 4 02 The programming screen closes automatically after 60 seconds witho...

Page 33: ...nge Time Press the time to open the keyboard Time is in 00h 24h format Change Date Press the date to open the keyboard Date is in dd mm yy format The validation of the changes is done only when you le...

Page 34: ...low the following steps For security reasons the number layout is done randomly and changes every opening of the password keyboard In case you want to remove the password lock from the User or Install...

Page 35: ...n protect Check or uncheck the box to change its state When the box is checked the screen protection is active and you can change the delay and the backlight level Change the activation time of the sc...

Page 36: ...Choice of Chlorine dosing relay You can either use the 220V P4 self powered relay to drive the dosing pump power supply or the P2 CRT relay to drive the pause contact Enable power If you have chosen t...

Page 37: ...count when is state change Change the contact activation delay Press the button to open the numeric entry keypad Change the relay configuration If you want to use the third relay output either alarm o...

Page 38: ...valid the time slot is displayed in green The hours must be entered in 00h 24h format f Temperature Pressing the Temperature button opens the programming window Temperature sensor By checking this box...

Page 39: ...15200 Change MODBUS parity Use the buttons on either side of the selection box to scroll through the different parities in one direction or the other None Even Odd Stop bit s information This part is...

Page 40: ...the button to open the keyboard and enter the address of the mysyclope server The maximum size is 30 characters Change TCP port Press the button to open the numeric keypad and enter the TCP port of th...

Page 41: ...nput box to open the numeric keypad and enter the IP provided by your IT manager Mask Gateway Same as previous DNS Automatic If the local Ethernet network on which the controller is connected is autom...

Page 42: ...the arrows to change the network mode Infra Network on which multiple elements can connect Ad Hoc Network on which only the regulator will be connected Security Press the arrows to select the security...

Page 43: ...s measurement value in mA the value of Chlorine or Bromine corresponding to the adjusted measurement of the calibration can be read The Offset and Gain information correspond to the calibration of the...

Page 44: ...he Modem module answer Error No answer from the modem check the hardware connection of the module PIN requested A PIN is needed PUK requested A PUK is needed PIN2 requested A PIN2 is needed PUK2 reque...

Page 45: ...from the modem check the hardware connection of the module xxx xxx xxx xxx Current IP of your controller a value other than 0 0 0 0 is correct WIFI Modem State messages Searching Waiting for the Modem...

Page 46: ...Initialization button When you press the button a confirmation window opens to confirm your choice WARNING all settings and configurations will be reset to factory Version Yes button Confirm the reset...

Page 47: ...must choose the channel you want to configure by pressing the left button for the pH channel and right for the Chlorine Bromine channel The programming menu of the chosen channel opens here pH gives a...

Page 48: ...ected on the mysyclope site you can choose not to stop the dosing and to send you an alarm email from the site Stop dosing option is not displayed if Not used mode is selected a Alarms Pressing the Al...

Page 49: ...point the previously activated actuator remains on Enter the hysteresis or proportional value Press the enter button to open the numeric keypad and enter the new one the new value When the dosing dir...

Page 50: ...rogramming window Change dosing direction Here you can choose the dosing direction depending on the type of treatment and the type of product injected Maximum dosing time This button is used to enter...

Page 51: ...ons are determining operations for good working of the controller and best treatment of the swimming pool A bad calibration could be dangerous for your health and your swimming pool It can cause corro...

Page 52: ...rn to the factory version use the Clear tab Offset or pH7 calibration Select the Offset tab Press the key to open the input keyboard of the new near pH 7 Press calibrate to validate the entry In case...

Page 53: ...ating the chlorine measurement of the CU PT probe make sure that The pH is stabilized at its nominal operating value of your installation The Chlorine or Bromine is also stabilized at the nominal oper...

Page 54: ...red Reset to factory calibration Select the Erase tab Press Clear Calibration OK information is displayed in green to confirm the deletion Warning this operation is irreversible Once confirmed by the...

Page 55: ...PE Electronique S A S Z I A ropole Pyr n es 64 230 SAUVAGNON Tel 33 05 59 33 70 36 Fax 33 05 59 33 70 37 Email service technique syclope fr 2019 2020 by SYCLOPE Electronique S A S Subject to modificat...

Reviews: