Technical characteristics and functions
Page 11/55
SYCLOPE TERE’O Touch XL
®
Installation, starting, and programming instructions
III.
Technical characteristics and functions
1)
Technical characteristics
Main features
Type
Specification(s)
Marker(s)
Consumption
12 W Max (Without connected dosing accessories)
-
Power supply requirements
90-240VAC 50/60Hz
-
Electrical protection
Fuse 160 mA. Reset by power interruption
F4
Operating temperature (°C) -5 °C to 45 °C (23 °F to 113 °F)
-
Case material
ABS or Polycarbonate (USA and Canada)
-
Dimensions of the case
Length: 213 mm (8.4 inches)
Width: 185 mm (7.3 inches)
Height: 118 mm (4.6 inches)
-
Case weight
1.1 kg
-
Display
4.3-inch colour LCD screen Resistive touch
-
Environment
Storage temperature
-10 °C to 70 °C (10 °F to 158 °F)
-
Humidity
Max. 90% without condensation
-
Protection rating
IP 65
-
Product certification
CE
-
Electromagnetic
compatibility
Class B disruption tests comply with EN61326-1
Class B disruption tests comply with EN61326-2-6
Class B disruption tests comply with EN55011
Harmonics tests comply with EN61000-3-2
Harmonics tests comply with EN61000-3-3
Immunity tests comply with EN61000-4-2
Immunity tests comply with EN61000-4-3
Immunity test EN61000-4-4
Immunity tests comply with EN61000-4-5
Immunity tests comply with EN61000-4-6
Immunity tests comply with EN61000-4-8
Immunity tests comply with EN61000-4-11
-
Standard
EN 61000 Electromagnetic compatibility (CEM)
EN 61326 Electrical measuring, control and laboratory equipment
for a standard environment (class B home use)
-
Entrées
Measurement inputs
1 potentiometric input (pH)
1 4…20mA isolated input Tension 24VDC (temperature)
1 4…20mA isolated input Tension 24VDC (chlorine/Bromine)
POT
IN1
IN2
Level inputs
1 Tank level contact input (pH)
1 Tank level contact input (Chlorine/Bromine)
K1
K2
Water circulation input
1 water circulation contact input
Aux
Sorties
Outputs relay
2 power relay output
-
Max. 3.15A / 250 VAC
-
Transient current Max 15A <1ms
2 ON/OFF relay output
P3 et P4
P1 et P2
REF
Reference 24V output for power supply to external measurement
converters or flow sensors.
VREF
Reference 12V output.
PWR
Port de communication
RS485
1 RS485 communication port compatible protocol type Modbus RTU
Protection des sorties de dosage
Internal fuse
Glass fuse 5x20mm 3.15A 250V time delay
FUSE1