A-96.250.211 / 171121
AMI Sodium P
Program List and Explanations
94
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active, acknowl
-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Message List
1.2.1
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then the
oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation
:
View the designation of instrument.
Version
: Firmware of instrument (e.g. V6.22-06/21).
2.1.3
Factory Test
: Test date of the instrument, motherboard and frontend
QC factory test.
2.1.4
Operating Time
: Years/days/hours/minutes/seconds.
2.2 Sensors
2.2.1
Sodium Sensor:
Current value
: Shows the sodium concentration in ppm.
(Raw value)
: Shows the uncompensated potential in mV.
2.2.1.5
Cal. History:
Shows the diagnostic values of the last calibrations of
the sodium electrode.Max. 65 data records are memorized.
2.2.2
pH Electrode:
Current Value:
Shows the measured pH value.
(Raw value):
Shows the electrode voltage in mV.
Typical offset of sodium sensor
125 mV
Max. tolerated offset
20
mV
Typical slope of sodium sensor
59
mV/decade Na
Max. limits
3
mV