A-96.250.211 / 171121
AMI Sodium P
Maintenance
78
3
Mount electrode and wait 2-3
minutes.
4
Screw standard bottle 1 onto the
holder and swing bottle upwards.
5
Wait until the process has finished.
6
Press <Enter> to save.
7
Remove the bottle from the bottle
holder.
8
Press <Enter> for 2-point-calibration
or <Exit> if you only want to make a
one-point calibration.
Standard Sodium
and wait
<Enter> to continue
3.1.1.5
Mount electrode
2 - 3 minutes
Standard Sodium
onto holder and swing
<Enter> to continue
3.1.1.5
Screw standard 1 bottle
bottle upwards
Standard Sodium
Current Value
199 ppb
<Enter> to save
3.1.1.5
Standard 1
200 ppb
Offset
405 mV
Standard Sodium
<Exit> if only one
<Enter> to continue
3.1.1.5
Remove bottle. Press
cal. point is needed