A-96.250.211 / 171121
AMI Sodium P
Troubleshooting
84
7.
Troubleshooting
7.1. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is
exceeded.
Such Errors are marked
E0xx
(bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re
-
covered (i.e. Sample Flow low).
Such Errors are marked
E0xx
(
bold and orange
)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked
E0xx
(
bold and red
)
Error or
fatal Error
Error not yet acknowledged.
Check
Pending Errors 1.1.5
and take
corrective action.
Navigate to <Messages>/<Pending
Errors>.
Press [ENTER] to acknowledge the
Pending Errors.
The Error is reset and saved in
the Message List.
23 B/s
HOLD
PO4
14:10:45
R1
0.5 ppm
R2
1.1
Pending Errors
Messages
Message List
1.1.5
Pending Errors
Error Code
E002
Alarm low
<Enter> to Acknowledge