AMI Sodium P
Maintenance
73
Disassemble
the flow cell
The flow cell can be easily disassembled. Proceed as follows:
1
Shut off the instrument.
2
Stop sample flow.
3
Drain the constant head completely.
4
Remove the constant head cover [A].
5
Pull the constant head tube [B] out of the flow cell block [F].
6
Release the set screw [G] with a 1 mm allen key.
7
Unscrew and remove the fixing screw [E].
8
Remove the standard bottle holder [D] from the flow cell block.
9
Clean all acrylic parts with a soft brush (bottle cleaner) and soapy
water.
Note:
Grease the O-rings only with teflon paste or spray.
Assemble the
flow cell
1
Replace all O-rings [C] before assembling the flow cell.
CAUTION
Damage of flow cell block possible.
If screws which are screwed into the flow cell block are tightened
too much, the thread or the flow cell block may be damaged.
Only tighten such screws slightly.
2
Push the standard bottle holder [D] into the flow cell block.
3
Slightly tighten the fixing screw [E].
4
Lock the fixing screw with the set screw.
5
Push the constant head tube [B] into the flow cell block [F].
6
Put the cover onto the constant head tube.