Maintenance
6.3.
Maintenance of the Sensor
6.3.1
Remove the Sensor form the Flow Cell
To remove the sensor form the flow cell proceed as follows:
1
Press the locking pin [G] down.
2
Turn the locking screw [H] with a 5 mm allen key counterclock-
wise 180°.
The locking pin remains down.
3
Remove the sensor.
Cleaning
If the sensor is slightly contaminated, clean it with soapy water and
a pipe cleaner. If the sensor is strongly contaminated, dip the tip of
the sensor into 5% hydrochloric acid for a short time.
6.3.2
Install the Sensor into the Flow Cell
1
Make sure that the locking mechanism is in unlocked position
(locking pin in position [G] and security screw in position [H]).
2
Put the sensor into the flow cell with the alignment marks [E] in
line.
3
Turn the locking screw with a 5 mm allen key clockwise 180°.
The locking pin moves up in lock position.