22
A-96.250.341 / 061015
AMI Powercon
Installation
3.
Installation
3.1.
Installation Checklist Monitors
Check
Instrument’s specification must conform to your AC power ratings.
Do not turn on power until instructed to do so.
On site require-
ments
100–240 VAC (± 10%), 50/60 Hz (± 5%) or 24 VDC, isolated
(±10%) power outlet with ground connection and 30 VA
Sample line with sufficient sample flow and pressure (see
).
Installation
Mount the instrument in vertical position.
Display should be at eye level.
Connect sample inlet and outlet.
Monitor: Sensors are already mounted.
Single flow cell: Mount sensors (see
, and connect cables (see
).
Electrical wiring
Connect all external devices like limit switches and current loops
see
.
Connect power cord; do not switch on power yet!
Cation
exchanger
Only for AMI Powercon acid!
Fill up cation exchanger bottle with high purity water.
Remove the empty bottle and install the cation exchanger bottle.
With pre-rinse set-up, install a cation exchanger bottle to the
second flange.
Power-up
Open sample flow and wait until flow cell is completely filled.
Switch on power.
Adjust sample flow.
Instrument
setup
Program all sensor parameters see
Program the required temperature compensation.
Program all parameters for external devices (interface,
recorders, etc.).
Program all parameters for instrument operation (limits, alarms).
Run-in period
If the conductivity value of the sample is very low, the sensor might
need some time until the correct reading is displayed