66
A-96.250.871 / 120417
AMI CACE
Program List and Explanations
5.3.2 and 5.3.3
Relay 1 and 2:
The contacts can be set as normally open or nor-
mally closed with a jumper. See
The function of relay contacts 1 or 2 is defined by the user.
NOTICE:
The navigation in the menu <Relay 1> and <Relay 2>
is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1
First select the functions as:
– Limit upper/lower
– Control upwards/downwards
– Timer
– Fieldbus
2
Enter the necessary data depending on the selected function.
The same values can also be entered in menu
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter:
select a process value
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Cond. 1 (sc)
0–3000
S
Cond. 2 (cc)
0–3000
S
Temp. 1
-25 to +270 °C
Temp. 2
-25 to +270 °C
Difference
0–3000
S
Sample flow
0–20 l/h
pH
0–14 pH
Ammonia
0–500 ppm