Instrument Setup
4.
Instrument Setup
After the analyzer is installed according to the previous instructions,
connect the power cord. Do not switch on power, yet!
4.1.
Establish sample flow
1
Open the sample tap
2
Check inlet pressure
3
Wait until the system has been completely filled
4
Switch on power
5
Let the instrument run in for 1 h
4.2.
Programming
Sensor
parameters
Program all sensor parameters in Menu Installation-Sensors:
menu 5.1.2.1.1 for sensor 1 and
menu 5.1.2.2.1 for sensor 2.
The sensor characteristics are printed on the label of each sensor.
Enter for each sensor separately the:
Cell constant [cm
-1
]
Temperature correction [°C]
Cable length. If the flow cell is installed on the monitor, set the
cable length to 0.0 m.
Temperature compensation: The default setting for sensor 1
(specific conductivity) is ammonia. For sensor 2 (cation
conductivity), the default setting is strong acids.
87-344.203
UP-Con1000SL
Sensor type
SW-xx-xx-xx
ZK = 0.0417
Cell constant
SWAN AG
DT = 0.06 °C
Temperature correction