![Swann AMI CACE Manual Download Page 25](http://html1.mh-extra.com/html/swann/ami-cace/ami-cace_manual_1411469025.webp)
AMI CACE
Installation
A-96.250.871 / 120417
23
3.6.3
Relay 1 and 2
NOTICE:
Max. load: 1 A/250 VAC
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a re-
lay as normally closed, set the jumper in the upper position.
NOTICE:
Some error codes and the instrument status may
influence the status of the relays described below.
For programming see
, Menu Installation
Relay
config.
Terminals
Jumper
pos.
Description
Relay configuration
Normally
Open
6/7: Relay 1
8/9: Relay 2
Inactive (opened) during
normal operation and
loss of power.
Active (closed) when a
programmed function is
executed.
Normally
Closed
6/7: Relay 1
8/9: Relay 2
Inactive (closed) during
normal operation and
loss of power.
Active (opened) when a
programmed function is
executed.
6
0V
7
6
0V
7
A
B
Jumper set as normally open (standard setting)
Jumper set as normally closed
A
B
Summary of Contents for AMI CACE
Page 1: ...AMI CACE Version 6 20 and higher A 96 250 871 120417 Operator s Manual...
Page 79: ...AMI CACE Index A 96 250 871 120417 77 Terminals 20 22 23 26 U USB Interface 27 W Wire 18...
Page 80: ...78 A 96 250 871 120417 AMI CACE Notes 12 Notes...