AMI CACE
Installation
A-96.250.871 / 120417
15
3.
Installation
3.1.
Installation Checklist Monitors
Check
Instrument’s specification must conform to your AC power ratings.
Do not turn on power until instructed to do so.
On site require-
ments
100–240 VAC (± 10%), 50/60 Hz (± 5%) or 24 VDC, isolated
(±10%) power outlet with ground connection and 30 VA
For sample requirements see
Instrument Specification, p. 12
).
Installation
Mount the instrument in vertical position.
Display should be at eye level.
Remove the end caps from tubes 1, 2, 3, 5 and 10 and connect
the tubes according to
.
Connect sample inlet and outlet.
Electrical Wiring
Connect all external devices like limit switches, current loops and
pumps (see
).
Connect power cord; do not switch on power yet!
Power-up
Open sample flow and wait until the instrument is completely
filled.
Check inlet pressure.
Switch on power.
Instrument
set-up
Program all sensor parameters (see
).
If required activate calculations (see
).
Program all parameters for external devices (interface,
recorders, etc.).
Program all parameters for instrument operation (limits, alarms).
Program display screens.
Run-in period
Let the instrument run continuously for 1 h.