Troubleshooting
7.3.
Replacing the EDI module
Select Menu 3.3 (Maintenance/Exchange EDI) and follow the in-
structions on the display.
Status of relays and signal outputs during the procedure:
Signal outputs are on hold
All limits are switched off
Overview
To unmount the EDI module, unscrew screws [A] and [D] and the
upper ends of tubes [1], [2] and [3].
Totalizers
and date of
exchange
A
B
Top screws (2x)
Holder
C
D
EDI module
Bottom screw
C
C
D
A
B
B
At the end of the procedure, the user
is asked if the EDI module has been
exchanged.
If <yes> is selected, the totalizers in
the diagnostics menu are reset and
the date of exchange is saved.
3.3.1
Exchange EDI
Has the EDI module
been exchanged?
yes
no