Troubleshooting
E038
Temp. 2 Alarm low
– check process
– check programmed value, see
E043
EDI out of range
– check sample inlet pressure and
acknowledge this error message
– If the problem persists, stop sample
flow and call service
E044
No sample flow
– check sample inlet pressure.
– check if the following components are
clogged:
inlet filter (if installed)
tubes
EDI module
– If necessary, replace clogged parts.
See
Replacing the EDI module, p. 44
.
E045
EDI DAC disconnected
– Stop sample flow and call service
E046
EDI ADC disconnected
– Stop sample flow and call service
E049
Power-on
– none, normal status
E050
Power-down
– none, normal status
E065
EDI module exhausted
– replace EDI module, see
.
Error
Description
Corrective action