AMI Turbiwell
Safety Instructions
9
1.3. Restrictions for use
Sample Re
-
quirements
Flow rate: 20–60 l/h
Temperature: 1–45 °C.
Swan recommends that the sample temperature is no
more than 20 °C above the ambient temperature.
The outlet has to be pressure-free against atmosphere.
Sample degasser: Only to be used for turbidity below 1 FNU/
NTU.
Flow rate of sample inlet at degasser: 10–12 l/h.
Maximum turbidity of the sample:
– Turbiwell 7027 and Power: 200 FNU
– Turbiwell W/LED: 100 NTU
CAUTION
Wrong measured values due to dirty optical components.
Touching the optical components in the cover of the measuring
cell may result in wrong measured values.
Cleaning and re-calibration at SWAN is necessary.
Never touch the optical components.