A-96.250.511 / 091120
AMI Turbiwell
Product Description
13
If the optional sample degasser is used, the sample first flows
through this device before entering the measuring chamber.
The drain valve [K] is used to empty the measuring chamber for
maintenance work like cleaning the measuring chamber or perform
-
ing a verification.
Verification
For the verification, three different verification kits (low turbidity, high
turbidity, liquid) can be used. The kits are available as an option.
Calibration
The AMI Turbiwell is factory-calibrated, therefore it is not necessary
to perform a calibration in the field. The emission intensity of the LED
is monitored by an external photodiode. A loss of intensity due to ag
-
ing will be automatically compensated.
To meet the requirements of some public authorities it is possible to
calibrate the instrument with a formazine standard.