A-96.250.511 / 091120
AMI Turbiwell
Program List and Explanations
101
5.3.2.342.2
Monday
: Possible settings, on or off
to
5.3.2.342.8
Sunday
: Possible settings, on or off
5.3.2.44
Run Time
: see Interval
5.3.2.54
Delay
: see Interval
5.3.2.6
Signal Outputs
: see Interval
5.3.2.7
Output/Control
: see Interval
The relay will be switched via the Profibus input. No further parame
-
ters are needed.
5.3.4
Input:
The functions of the relays and signal outputs can be defined
depending on the position of the input contact, i.e. no function,
closed or open.
5.3.4.1
Active:
Define when the input should be active:
5.3.4.2
Signal Outputs:
Select the operation mode of the signal outputs
when the relay is active:
5.3.4.3
Output/Control:
(relay or signal output):
5.3.4.4
Fault:
5.3.2.1
Function = Fieldbus
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Continuous:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 [mA] respectively. Errors, except fatal
errors, are not issued.
Continuous:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the message
list. The Alarm relay closes when input is active.