AMI Turbiwell
Maintenance
69
Verification
Kit Turbiwell
Liquid
The Verification Kit Turbiwell Liquid can be filled with any aqueous
suspension or emulsion, provided that the suspension or emulsion is
compatible with the materials used.
Note:
The Verification Kit Turbiwell Liquid is suitable for
verification, but not for calibration!
WARNING
Please observe the necessary safety precautions when
handling hazardous chemicals.
Read the Safety Data Sheets carefully!
Filling
To fill the Verification Kit Turbiwell Liquid, proceed as follows:
1
Unscrew both screw plugs [A].
2
Fill 13.5 to 14 g of the standard into the cuvette through one of
the openings.
Note:
The cuvette must not be completely filled with liquid,
otherwise it may be damaged by overpressure if the temperature
rises.
3
Close the cuvette with the two screw plugs.
4
Perform the steps specified in
.
5
Perform the matching procedure, see
Housing:
PET
Windows:
PMMA
Screw plugs:
PVDF
A
B
C
Screw plugs
Cuvette
Window (light beam in)
D
E
Window (light beam out)
Window (light beam out)
A
B
D
C
E