A-96.250.511 / 091120
AMI Turbiwell
Installation
32
Electrical
Connection
WARNING
Electrical shock hazard
Before opening the AMI Transmitter switch power off.
17
Use one of the PG7 cable glands to feed the cable of the sensor
into the AMI transmitter housing.
18
Connect the cable to the terminals according to the
Exchanging
the capillary
The capillary determines the flow resistance and the maximum flow
rate. With the pressure regulator at the inlet side of the capillary the
flow rate can be adjusted within the given range, see diagram below.
The standard capillary is an FEP tube with an inner diameter of
1 mm and a length of 500 mm. If you have a low sample pressure or
you need a high sample flow, a shorter capillary with 186 mm length
can be used.
0
0
5
10
15
20
25
30
35
40
0.5
1
bar
l/h
1.5
2
2.5
3
3.5
186
500