TRIO MPC-165 SERIES MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.67K2 (20201130) (FW V2.62)
26
Figure 3-13. Changing the MP-865/M micropipette/headstage mount and angle.
To change the headstage/micropipette, loosen the screw in the center of the rotary dovetail
clamp. Slide the headstage upward out of the dovetail groove. Make any adjustments needed
of the headstage, and then tighten down (but do not over tighten) the lock screw in the
center of the holding bracket.
Normal position
Partially swung
Fully swung
Rotary dovetail
angle lock set
screw
Rotary dovetail
clamp screw
Rotary micropipette/-
headstage exchange
swing-out lock/release