TRIO MPC-165 SERIES MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.67K2 (20201130) (FW V2.62)
3
DISCLAIMER
The TRIO MPC-165 series system consists of one or two electromechanical
micromanipulator devices and one TRIO MPC-100 ROE (Rotary Optical Encoder) with
integrated controller. The purpose of the system is for the manipulation at the micro level of
micropipettes and probes used in conjunction with a microscope. No other use is
recommended.
This instrument is designed for use in a laboratory environment. It is not intended nor
should it be used in human experimentation or applied to humans in any way. This is not a
medical device.
Unless otherwise indicated in this manual or by Sutter Instrument Technical Support for
reconfiguration, do not open or attempt to repair the instrument.
Do not allow an unauthorized and/or untrained operative to use this device.
Any misuse will be the sole responsibility of the user/owner and Sutter Instrument Company
assumes no implied or inferred liability for direct or consequential damages from this
instrument if it is operated or used in any way other than for which it is designed.
SAFETY WARNINGS AND PRECAUTIONS
Electrical
Operate the TRIO MPC-100 using 110 – 240 VAC., 50-60 Hz line voltage. This instrument
is designed for use in a laboratory environment that has low electrical noise and
mechanical vibration. Surge suppression is always recommended
NOTE: There are no user-replaceable fuses in the TRIO MPC-100 system.
The TRIO MPC-100 system’s power supply consists of an external AC to DC
switching power adapter. If the external power adapter is damaged due to a mains over or
under voltage, it must be replaced.
GROUNDING/EARTHING: Proper grounding protects the ROE/controller
electronics, reduces/eliminates electromagnetic interference, and improves the safety of the
system operator. The ROE/controller provides a socket (labeled GROUND) that accepts a
banana plug attached to a suitably gauged insulated wire, the other end of which (alligator
clip) connects to a solid, proper ground.
Avoiding Electrical Shock and Fire-related Injury
Always use the grounded power cord provided to connect the system’s power adapter to a
grounded/earthed mains outlet (3-prong). This is required to protect you from injury if an
electrical hazard occurs.
Do not disassemble the system. Refer servicing to qualified personnel.
To prevent fire or shock hazard do not expose the unit to rain or moisture.
Electromagnetic Interference
To comply with FDA and CE/EU electromagnetic immunity and interference standards; and
to reduce the electromagnetic coupling between this and other equipment in your lab always
use the type and length of interconnect cables provided for interconnecting the electro-
mechanical devices and ROE/controller (refer to Technical Specifications for more details).