Upon deciding to directly install the TRIO MPC-100 system in your rig, it is useful to follow
the initial setup procedure to learn how to move the units to allow easy access to the
mounting screws.
1.
With the power switch on the back of the ROE in the OFF (0) position, connect the power
adapter’s 24VDC cable to the POWER receptacle.
Figure 2-6. Rear of TRIO MPC-100 ROE/Controller cabinet
2.
With the power OFF (rear panel switch in the “0” position), connect a well-
grounded/earthed wire to the GROUND banana plug receptacle.
3.
With the power OFF, connect the male end of the DB -25 cable to either MANIPULATOR
connector on the ROE, the other end of which is connected to the MP-865/M
micromanipulator electromechanical stage. (See cautionary note below.)
4.
Verify that the ten switches associated with the appropriate MANIPULATOR (A or B) on
the rear of the ROE are set as desired.
5.
Power up the system by moving the power switch on the rear of the ROE to the “1”
position.
*
CAUTION: NEVER CONNECT OR DISCONNECT THE ROE/CONTROLLER
FROM EITHER ELECTROMECHANICAL MANIPULATOR/STAGE WHILE THE POWER
IS ON!
2.7
ROE/Controller Rear Panel Controls and Configuration
2.7.1
Power Switch
The power switch for the TRIO MPC-100 system is located on the rear panel of the
ROE/controller. At power up, the microprocessor in the ROE/controller scans the attached
equipment and configures the system accordingly.
CAUTION: When changing the configuration switches described in the following paragraphs,
make sure that controller’s power switch is turned off.
Power
Switch
USB
Device
Receptacle
Ground/Earth
Post
Manipulator A
cable connector
1
0
Power
Input
Socket
Manipulator B
configuration
switches
Manipulator B
cable connector
Manipulator A
configuration
switches