Description
12
Fuses:
Fuse
HC-4A
HC-4A2
Circuit
Location
FU1
0.63 AT
1.0 AT
Cold Head
Front Panel
FU2
0.63 AT
1.0 AT
Cold Head
Front Panel
FU3
5.0 AT
5.0 AT
Cold Head
Front Panel
FU4
5.0 AT
5.0 AT
Cold Head
Front Panel
FU5
0.4 AT
0.4 AT
Control
Inside Electrical Chassis
FU6
0.4 AT
0.4 AT
Control
Inside Electrical Chassis
FU7
1.0 AT
1.0 AT
Fan and Circulator
Inside Fan Cabinet
FU8
1.0 AT
1.0 AT
Fan and Circulator
Inside Fan Cabinet
FU9
0.63 AT
0.63 AT
Fan and Circulator
Inside Fan Cabinet
Gas Supply and Return Couplings:
Both are self-sealing, size 8, male, Aeroquip, bulkhead
couplings and are the points of connection on the front panel for the rest of the system.
Power On/Off Switch:
This on/off rocker switch starts and stops the compressor. The switch
lights to indicate that power is on to the compressor.
Pressure Gauge:
A pressure gauge indicates gas pressure in the supply line. When the
compressor is not running, the gauge shows the equalization pressure.
Summary of Contents for HC-4A
Page 2: ......
Page 6: ...iv This page is intentionally blank ...
Page 10: ...4 This page is intentionally blank ...
Page 12: ...6 This page is intentionally blank ...
Page 14: ...8 This page is intentionally blank ...
Page 36: ...30 This page is intentionally blank ...
Page 38: ...32 This page is intentionally blank ...
Page 59: ...Troubleshooting 53 Figure 8 HC 4A Wiring Diagram Main Electrical Chassis and Components ...
Page 60: ...Troubleshooting 54 Figure 9 HC 4A2 Wiring Diagram Main Electrical Chassis and Components ...
Page 61: ...Troubleshooting 55 Figure 10 HC 4A and HC 4A2 Wiring Diagram Air Cooled Unit ...
Page 62: ...Troubleshooting 56 Figure 11 HC 4A Electrical Schematic ...
Page 64: ...Troubleshooting 58 Figure 13 HC 4A2 Electrical Schematic Air Cooled Unit ...
Page 68: ...Parts 62 Figure 15 Parts Identification for HC 4A 12 14 19 23 ...
Page 69: ...Parts 63 Figure 16 Parts Identification for HC 4A2 12 14 19 23 ...
Page 70: ...Parts 64 Figure 17 Parts Identification 29 28 30 31 ...